Bergantung kepada jenis peranti, laporan hendaklah dikemukakan mengikut standard yang disenaraikan di bawah : | Depending on the type of devices, the reports shall be submitted as per standards listed below :

  • ISO 594-1: Conical Fittings with 6% (Lure) Taper for Syringes, Needles and Certain Other Medical Equipment - Part 1: General Specifications
  • ISO 594-2: Conical Fittings with 6% (Lure) Taper for Syringes, Needles and Certain Other Medical Equipment - Part 1: Lock Fittings
  • ISO 720: 1985: Glass - Hydrolytic Resistance of Glass Grains a 121º C - Method of Test and Classification
  • ISO 7864: Sterile Hypodermic Needles for Single Use
  • ISO 7886-1: Sterile Hypodermic Syringes for Single Use - Part 1: Syringes for Manual Use
  • ISO 7886-2: First Edition 1996-05-15: Sterile Hypodermic syringes for Syringes for Single Use - Part 2: Syringes for use with power driven syringe pumps
  • ISO 7886-3: First Edition 1996-05-15: Sterile Hypodermic syringes for Syringes for Single Use - Part 3: Auto-disposable syringes for fixed-dose immunization
  • ISO 11040-4: Prefilled Syringes Part 4: Glass Barrel for Injectables
  • ISO 10993-1: Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
  • ISO 11608-1: Pen-Injectors for Medical Use – Part 1: Pen-injectors – Requirements and test methods
  • ISO 11608-2: Pen-Injectors for Medical Use – Part 2: Needles – Requirements and test methods
  • ISO 11608-3: Pen-Injectors for Medical Use – Part 3: Finished Cartridge – Requirements and test methods
  • ISO 11608-4: Pen-Injectors for Medical Use – Part 4: Requirements and test methods for electronic and electromechanical pen-injectors
  • ISO 21649: Needle-Free Injectors for Medical Use - Requirements and Test Methods.
  • ASTM D4169:, Standard Practice for Performance Testing of Shipping Containers and Systems
  • ANSI/AAMI/ISO 17665-1: 2006, Sterilization of Health Care Products – Moist Heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices
  • ANSI/AAMI/ISO 11135-1:2007,Sterilization of Health Care Products – Ethylene oxide – Part 1: Requirements for development, validation, and routine control of a sterilization process for medical device
  • USP 27:2004, Sterility, Biocompatibility, Biological Tests and Assays, Bacterial Endotoxin Test (LAL), Pyrogen Test (USP Rabbit Test), or other applicable tests related to the drug/biological product and delivery of the drug/biological product
  • AAMI/ANSI/ISO 11737-1:2006, Sterilization of medical devices-microbiological methods-Part 1: Determination of the population of microorganisms on product
  • ANSI/AAMI/ISO 11607:2006, Packaging for terminally sterilized medical devices The report of all tests mentioned above must include the objectives, methodology, results, analysis and manufacturer's conclusions.
Untuk pertanyaan lanjut, sila hubungi 
BAHAGIAN KAWALAN PRA PASARAN (BKPP)
03 - 8230 0376

For further inquiries, please contact
PRE-MARKET CONTROL DIVISION (BKPP)
03 - 8230 0376